The
GAR-B2 reagent gives the brightest bioluminescent signal for
both Gaussia and renilla luciferase assays
Dilute
100X coelenterazine substrate with assay dilution buffer
supplied with each product just before use.
Assay 20
µl of the gaussia luciferase sample with 50 µl Gaussia
luciferase assay reagent (GAR-B2). If activity is too high
dilute with serum-free DMEM but use 20 µl total volume per assay
using 50 µl of Gaussia luciferase assay reagent. This assay
reagent is compatible with measuring renilla and Gaussia
luciferase activities in cells lysed with either lysis buffers
from Targeting Systems or the Promega passive lysis buffer.
Mix well
and read in the luminometer.
|